Reference mark signal
Oscilloscope display x - t
Change the trigger setting of the oscilloscope basic setting
as follows:
Trigger channel A
Manual triggering (AC or DC)
Trigger negative edge
Traverse the reference mark to be checked
oscillatorily (forward/backward).
At the oscilloscope turn the trigger
potentiometer to set the trigger threshold
(LEVEL) such that the reference mark
signal is depicted as stationary image
on the screen.
Note:
Schematic display of Ue1+2 on the oscilloscope.
The actual amplitude is higher than displayed.